Agilent 4155B/4156B User’s Guide Vol.2, Edition 5 5-37
Analyzing Measurement Results
Automatic Analysis
To Display Marker at Specified Point
1. Press Display front-panel key.
2. Select ANLYSIS SETUP primary softkey. The DISPLAY: ANALYSIS SETUP
screen is displayed.
3. Move pointer to field (1), then select secondary softkey to set desired data
variable name.
4. In field (2), enter desired expression.
5. In field (3), select:
• AFTER secondary softkey if you want to set a search start condition for
finding specified point.
• DISABLE secondary softkey to disable (clear) the AFTER settings.
6. If you selected AFTER in field (4), select secondary softkey to set desired data
variable.
7. If you selected AFTER in field (5), enter desired expression.
The marker can be displayed on the measurement curve only. So, if no measurement
data satisfy the specified condition, the nearest measurement point is used.
For the meaning of expression that you can enter in step 4 and 7, see “Expression”
in Chapter 7.
To specify a point between two measurement points
Set Interpolate field to ON.