3-54 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Measurement Ranging Mode
Auto Ranging
The monitor unit automatically searches for and measures at the range that provides
the highest resolution as follows:
V measurement
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 110 % of the range, then the unit performs the
measurement.
I measurement
• 1 A to 1 mA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
If the measurement value is less than 1 % of the present range and the present
range is 100 mA or higher range, the range changes down two ranges instead of
one range.
• 100 nA to 100 pA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
• 10 pA
The unit changes to the next higher range when the measurement value exceeds
104 % of the present range.