EasyManuals Logo

Agilent Technologies 4155B User Manual

Agilent Technologies 4155B
414 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
3-54 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Measurement Ranging Mode
Auto Ranging
The monitor unit automatically searches for and measures at the range that provides
the highest resolution as follows:
V measurement
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 110 % of the range, then the unit performs the
measurement.
I measurement
• 1 A to 1 mA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
If the measurement value is less than 1 % of the present range and the present
range is 100 mA or higher range, the range changes down two ranges instead of
one range.
• 100 nA to 100 pA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
• 10 pA
The unit changes to the next higher range when the measurement value exceeds
104 % of the present range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals