Agilent 4155B/4156B Users Guide Vol.2, Edition 5 4-31
Making a Measurement
Sweep Measurements
To Execute Calibration
To execute all self-calibration test items, perform the following:
1. Press System key in the PAGE CONTROL key group.
2. Select CALIB/DIAG primary softkey.
3. In the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS screen, select CALIB
ALL secondary softkey. After the calibration execution, each test result is
displayed in each STATUS field.
To set auto-calibration
Move the pointer to the AUTO CALIB field, then select the ON secondary softkey.
Self-calibration is performed automatically every 30 minutes.
Be aware that measurement cannot be performed while calibration is executed.
To Use Offset Cancel Function
1. Press Meas key in the PAGE CONTROL key group.
2. Select MEASURE SETUP primary softkey.
3. In the ZERO CANCEL field, select ZERO CANCEL ON/OFF secondary
softkey to toggle the zero offset cancel mode between on and off. Then ON or
OFF appear automatically in each unit field depending on the measurement
range.
4. Press green key, then Stop key to measure the zero offset data. Then this
measured data is used to compensate the measurement results when
measurement is performed.
For more details about the zero offset cancel function, refer to Zero Offset Cancel
in Chapter 3.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals