6-24 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Screen Organization
MEASURE Screen Group
STOP CONDITION
• ENABLE/DISABLE
This field defines whether the stop conditions are enabled. Cannot ENABLE if
INITIAL INTERVAL < 2 ms. In this field, select:
• ENABLE secondary softkey to enable the stop conditions.
• DISABLE secondary softkey to disable the stop conditions.
• ENABLE DELAY
This field sets the enable delay time. The stop condition is ignored for the
enable delay time after the sampling starts. The resolution of enable delay time
is the initial interval time.
• NAME
NAME field sets the variable name or user function name that you want to
monitor for the stop conditions. Allowable variable names and user function
names are shown in the secondary softkey area.
• THRESHOLD
In the THRESHOLD field, you set the threshold value.
• EVENT
In the EVENT field, you set the event type as follows:
Val> T h event occurs when NAME value is greater than THRESHOLD.
Val< T h event occurs when NAME value is less than THRESHOLD.
|Val|>|Th| event occurs when absolute NAME value is greater than absolute
THRESHOLD value.
|Val|<|Th| event occurs when absolute NAME value is less than absolute
THRESHOLD value.
• EVENT NO.
EVENT NO. specifies sampling to stop if event occurs EVENT NO. times.
EVENT NO. can be an integer from 1 to 200.