Agilent 4155B/4156B User’s Guide Vol.2, Edition 5 6-57
Screen Organization
GRAPH/LIST Screen Group
• VAR2 STEP softkey (displayed only if VAR2 is defined)
Select VAR2 STEP secondary softkey to change the step value of the secondary
sweep VAR2. The present step value is shown on this softkey and in the data
entry area. Then you can change the value.
• COMP softkey (displayed only if VAR2 is defined)
Select COMP secondary softkey to change the compliance and power
compliance values of the secondary sweep VAR2. The present V or I
compliance value is shown on the middle line of this softkey, and the present
power compliance value is shown on the last line of this softkey.
Selecting COMP softkey highlights this softkey, and the present I or V
compliance value appears in the data entry area. Then you can change the value.
Then selecting COMP softkey again displays the power compliance value in the
data entry area. You can change the value. To disable the power compliance,
enter 0 (zero) or OFF.