3-22 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Knob Sweep Function
COMPLI-
ANCE
100.mA Sets compliance value for VAR1 channel. This softkey displays present
setting.
• setting
Press this softkey to display the compliance value in the data entry
area. Then, you can change the setting value by using the rotary
knob, numeric keys, or arrow keys in the Edit key group.
• setting range
Compliance range depends on measurement unit. See Chapter 1.
• default
VAR1 compliance value on the MEASURE: SWEEP SETUP
screen
HOLD
TIME
0.00s Sets hold time. This softkey displays present setting.
• setting
Press this softkey to display the hold time in the data entry area.
Then, you can change the setting value by using the rotary knob,
numeric keys, or arrow keys in the Edit key group.
• setting range : 0 to 655.35 s with 10 ms resolution
• default : hold time on the MEASURE: SWEEP SETUP screen
STEP
TIME
500us Sets step time which is the time width of each sweep step. This softkey
displays present setting.
• setting
Press this softkey to display the step time in the data entry area, then
you can change the setting value by using the rotary knob, numeric
keys, or arrow keys in the Edit key group.
• setting range : 500 ms to 100 ms with 100 ms resolution
• default : 500 ms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals