How to Use Laser Source Modules Laser Sources
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 113
How to Modulate the Optical Output
How to Change Modulation Source
The Modulation Source parameter, [Mod Src], allows you to modulate the
output signal. This parameter can be set to Off, Internal, Coherence
Control, or Backplane.
For details of the Off setting How to Disable Modulationon page 113.
For details of the Internal setting How to Use the Internal Modulationon
page 114
For details of the Coherence Control setting How to Increase Linewidth
on page 114
For details of the Backplane setting External Digital Modulation using
Input Trigger Connectoron page 147.
How to Disable Modulation
If <Off> is displayed as the Modulation Source parameter, [Mod Src], the
laser source outputs an unmodulated continuous-wave signal.
To disable modulation:
1 Move to the Laser Source channel and press [Details] softkey.
2 Move to the modulation source parameter, [Mod Src], and press Enter.
3 Move to <Off>, by using the cursor key, and press Enter.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals