How to Use Laser Source Modules Laser Sources
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 117
How to Use Triggers
Some Laser Source modules allow you to output trigger signals to external
measurement instruments.
How to Use Output Triggering
You can configure the output trigger connector to output a TTL signal at
the frequency of the internal modulation of a Agilent 8163A/B Series Laser
Source module.
To set your modules Output Trigger Configuration:
1 See How to Set the Trigger Configurationon page 69 for how to configure the
trigger connectors.
2 Move to the Laser Source channel and press [Menu].
3 Move to <Output Trigger Mode>, by using the cursor key, and press Enter. You
will see the screen in Figure 62 .
Figure 62 Output Trigger Mode
4 Move to one of the following, by using the cursor key:
<Disabled>, the output trigger mode is disabled.
•<Modulation>, the output trigger connector outputs a TTL signal at the frequen-
cy of the internal modulation. This signal is output whether the laser is switched
on or off.
5 Press Enter.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals