Error Messages Tuna ble Lase r Sources
396 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Selftests
These tests are started after initialization of the instrument. This test measures
defined parameters, and tests them to limits.
error message Main Peltier Temp. Selftest Failed (error value)
description Verifies whether the peltier temperature of the laser diode is within a given
range
error value 13010 Peltier temperature is not in target range
action Please return the unit to your local service center for repair
error message Main Thermo Temp. Selftest Failed (error value)
description Verifies whether the temperature of the thermopositioner is within a given
range
error value 13020 Thermopositioner is not in target temperature range
action Please return the unit to your local service center for repair
error message Main Peltier Curr. Selftest Failed (error value)
description Checks if peltier current for the laser diode peltier is below maximum
error value 13050 Laser peltier current is above maximum value
action Please return the unit to your local service center for repair
error message TCFS Temp. Selftest Failed (error value)
description Verifies if the peltier temperature is within target with a given uncertainty
error value 15010 TCFS peltier temperature is not in target range
action Please return the unit to your local service center for repair
error message TCFS Resistor Selftest Failed (error value)
description This test measures the R
d
of the photodiodes at channel A, B and M
error value 15021 Photodiode R
d
in channel A is below limit
15022 Photodiode R
d
in channel B is below limit
15023 Photodiode R
d
in channel C is below limit
action Please return the unit to your local service center for repair
error message TCFS Attenuator Selftest Failed (error value)
description Verifies whether an attenuator position is repeatable.
error value 15030 Attenuator could not be positioned
action Please return the unit to your local service center for repair