Tuna ble Laser S ou rces Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 395
error message Initialization Error (UpPiezoEnable, error value)
description Over current
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error - module temperature out of range
(please wait a little bit and try again)!
description Module temperature is too high
error value (no error value)
action Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.
error message Initialization Error - module temperature too high
(please wait a little bit and try again)!
description Module temperature is too high
error value (no error value)
action Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.