Fixed Laser Sources (Fabry Perot) Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 409
Fixed Laser Sources (Fabry
Perot)
error message E10010: Invalid Caldata
description The Caldata loaded on the module is invalid. Either the caldata
version number is not compatible with the firmware or the calda-
ta is corrupt (checksum incorrect)
test value (no test value)
action Please return the unit to your local service center for repair
error message E10012: Dual ported RAM write/read test failed
description Writing bit patterns and reading them back to/from the dual port-
ed ram failed
test value (no test value)
action Please return the unit to your local service center for repair
error message E10013: FPGA not correctly initialized
description Loading the FPGA with data failed. This indicates a defective
FPGA or an FPGA access problem
test value (no test value)
action Please return the unit to your local service center for repair
error message E10040: 10bit ADC test failed (a, b, c)
description The 10bit ADC has returned a value other than 512
test value a shows the ADC reading
b shows the lower limit
c shows the upper limit
action Please return the unit to your local service center for repair
error message E10050: Board temp out of limits (a)
description The board temperature is outside the allowed range
test value a measured board temperature
action Please return the unit to your local service center for repair