Errors appearing on pop-up menus Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 387
Errors appearing on pop-up
menus
Error on Module
error message Calibration data corrupted
description The calibration data loaded on the module is invalid. Either the caldata version
number is not compatible with the firmware or the caldata is corrupt (checksum
incorrect)
affected modules all modules
action Please return the unit to your local service center for repair
error message Laser safety error
description The laser control circuit has switched of the laser. Either the laser current or the
laser power was too high.
affected modules TLS
action Try to run the automatic realign from the menu or to operate the laser at a lower
power level. If the failure persists, please return the unit to your local service
center for repair.
error message Temperature out of range
description The temperature to operate the tunable laser source module was measured
outside given limits.
affected modules TLS
action Verify the fan is working and the ventilation system is not blocked. If the failure
persists, please return the unit to your local service center for repair.
error message Zeroing failed. Light?
description Zeroing has still detected some light. Either the input is not darkened properly
or there is some error on the powermeter module/head.
affected modules Powermeter
action Darken the optical input(s) properly. If the failure persists, please return the unit
to your local service center for repair.