Tuna ble Laser S ou rces Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 391
Tunable Laser Sources
Initialization Tests
These tests are executed every time the instrument is powered on, before the
selftests start.
error message Initialization Error (AttInitialize(), error value)
description Error during attenuator initialization. No ready message from attenua-
tor DSP after ...
error value 1 reset
2 init motor
4 find index
8 move to min position
action Please return the unit to your local service center for repair
error message Initialization Error (AttMovePos(zeroPos), error value)
description No ready message from attenuator DSP after moving to zero position
taken from caldata
error value 8 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (CalData CRC wrong: crc1, stored: crc2!)
description The checksum stored in the caldata does not match the calculated
checksum
error value crc1 calculated checksum
crc2 checksum read from caldata
action Please return the unit to your local service center for repair
error message Initialization Error (Calibration Data Invalid, error value)
description The calibration data was tagged as invalid before it was loaded on to
the module
error value 0 not significant
action Please return the unit to your local service center for repair
error message Initialization Error (CalOptZeroMeas, error value)
description Error in optical zero measurement
error value 1 not significant
action Please return the unit to your local service center for repair