Error Messages Tuna ble Lase r Sources
392 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
error message Initialization Error (DspMainCalcKFactor failed, error value)
description DSP K factor calibration failed
error value -1 no TCFS (according to caldata)
1 error when calibrating TCFS
2 DSP sweep failed
4 laser safety
action Please return the unit to your local service center for repair
error message Initialization Error (DspMainWriteCaldata(), error value)
description Error while writing cal data to DSP
error value 5 tried to access invalid array position
6 wrong dimension
7 unknown type found in variable array
8 wrong parameter when trying to access one of the DSPs cal
arrays
action Please return the unit to your local service center for repair
error message Initialization Error (Firmware CRC wrong: crc1, stored: crc2!)
description The checksum stored in the firmware does not match the calculated
checksum
error value crc1 calculated checksum
crc2 checksum read from caldata
action Reload the firmware of the module. If the failure persists, please return
the unit to your local service center for repair
error message Initialization Error (HwAccessInit, error value)
description Error setting DAC
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (InitDacs, error value)
description Error setting DAC
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (InstSettingInit, error value)
description Error creating semaphore
error value 2 (not significant)
action Please return the unit to your local service center for repair