Error Messages Tuna ble Lase r Sources
394 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
If one of the selftests fails, please try to reboot the instrument. If the selftest
continues to fail, please return the mainframe to your local service center for
repair.
error message Initialization Error (TASK_VERIFY, id1, id2)
description A task could not be started
error value id1 task name
id2 maximum available memory
action Reboot the instrument. If the failure persists, please return the unit to
your local service center for repair
error message Initialization Error (Tcfs Amplifier Calibration failed, error value)
description The calibration of the tuning control feedback system amplifier has
failed
error value 0 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TcfsAdpEEPROMInit, error value)
description Error reading tuning control feedback system adapter board EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TcfsEEPROMInit, error value)
description Error reading tuning control feedback system EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (Tcfs temperature not settled, error value)
description Temperature has not settled at the tuning control feedback system
error value 0 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TmpCalDataInit, error value)
description Error accessing caldata
error value 1 (not significant)