EasyManuals Logo

Agilent Technologies 8163B User Manual

Agilent Technologies 8163B
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #157 background imageLoading...
Page #157 background image
How to Use Auxiliary Functions Tunable Lasers
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 157
If Lambda Zero is recommended (Accuracy Class is not ACC1) the 10
th
bit
of the “questionable condition register is set (See the Agilent 8163A/B
Lightwave Multimeter, Agilent 8164A/B Lightwave Measurement System,
Agilent 8166A/B Lightwave Multichannel System Programming Guide).
This is done in both Auto Cal On and Off.
When Auto Cal is on, during sweeps, there is no interruption to the
measurement due to settling. Also no settling will occur during operation
at a fixed wavelength when at least one output is switched on. By
periodically checking the questionable condition register bit 10 you can
detect when Lambda Zero is recommended.
NOTE
NOTE
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals