Firmware Upgrade Process Firmware Upgrades
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 461
Agilent Module Firmware Upgrade Procedure
Agilent module firmware is upgraded via GPIB or a serial null-modem
cable.
You require a PC running MS Windows 95 or NT 4.0 or higher. To update via
GPIB the PC must be fitted with a suitable interface card. Alternatively, a
serial null-modem cable is required, such as is supplied with your Agilent
8163A/B mainframe.
You also require the module update program, which is supplied both on the
Support CD-ROM and as a web download. Copy Module_Update.exe to a
directory on your PCs hard drive, then execute Module_Update.exe to
install the module update program.
The readme.htm file associated with the module_update program contains
Notification of prerequisites,
Notification of known issues,
as well as detailed instructions for use.
Be sure to study this file before attempting to use the module update
software.
The floppy update procedure for the Agilent 81480A and
81640A/42A/80A/82A Tunable Laser Modules is no longer suppported.
Please update your modules via GPIB or a null-modem serial cable as
described above.
NOTE
NOTE
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals