Chapter 3 — Vector Network Analyzer 3-3 Calibration - 2-Port Calibration Procedure (OSLT)
MS20xxA VNA Master UG 10580-00166 Rev. C 3-7
12. Verify that calibration has been performed properly by checking that the Cal Status On
message is now displayed at the top of the status window.
Figure 3-2. 2-Port Calibration
OPEN
LOAD
SHORT
TEST PORT CABLE
(OPTIONAL)
TEST PORT CABLE
(OPTIONAL)
RF IN T
RF OUT TEST PORT
OPTIONAL ADAPTOR FOR
THROUGH CONNECTION
10580-00166C.book Page 7 Thursday, February 21, 2008 12:52 PM