2-6 Spectrum Analyzer Absolute Amplitude Accuracy Verification Spectrum Analyzer Verification
2-12 PN: 10580-00255 Rev. J MT8212E and MT8213E MM
10. Press the Span submenu key, use the keypad to enter 10, and press the kHz submenu key.
11. Press the Amplitude main menu and then press the Reference Level submenu key.
12. Use the keypad to enter 10 and press the dBm submenu key.
13. Press the Atten Lvl submenu key and enter 30, then press the dB submenu key.
14. Adjust the source power so that the power meter displays the corresponding desired Sensor B reading as
recorded for 0dBm in the Required Sensor B Reading column of Table A-5 on page A-5.
15. Press the Marker main menu and press the Peak Search submenu key.
16. Record the Marker 1 amplitude reading in the 0dBm row of Table A-6, “Spectrum Analyzer 50 MHz
Absolute Amplitude Accuracy” on page A-5.
17. Verify that the Marker 1 amplitude reading is within the specification.
18. Repeat Step 14 through Step 17 for the other power level settings. Refer to Table A-5 for Required
Sensor B Readings. Use Table A-6 to record test results. The last two settings are with the pre-amp
turned on, to ensure pre-amp functionality.