Spectrum Analyzer Verification 2-6 Spectrum Analyzer Absolute Amplitude Accuracy Verification
MT8212E and MT8213E MM PN: 10580-00255 Rev. J 2-13
Amplitude Accuracy Across Frequency Verification
This procedure is the second test that is used to verify the absolute amplitude accuracy of the Spectrum
Analyzer in the MT821xE Cell Master. The first procedure test was described in Section “50 MHz Amplitude
Accuracy Verification” on page 2-9.
Equipment Required
Anritsu MG3692X Synthesized Signal Source
Anritsu ML2438A Dual Channel Power Meter
Anritsu MA2442D High Accuracy Power Sensors (2)
10 MHz Reference Standard
Anritsu 34NN50A 50 ohm Adapter
Anritsu 34RKNF50 50 ohm Adapter
Anritsu 15NN50-1.5C RF Coaxial Cable
Aeroflex/Weinschel 1870A Power Splitter
Aeroflex/Weinschel 44-10 10 dB Fixed Attenuator