112
Z31900_00_01
Error Codes
103.310.107
117.310.126
Missing empty posi-
tion for security eject.
The cassette was searched for 
an empty position for executing 
a security eject attempt. But the 
cassette is completely filled.
• Activate Sample Down Detection 
Mode (Bruker service only).  
 
This allows a completely filled cas-
sette to be added to the SampleX-
press. Otherwise at least 1 empty 
position is needed in the cassette 
to execute a Security Eject of a 
potential sample in the magnet.
103.310.012
103.310.032
Missing empty posi-
tion.
The cassette was searched for 
an empty position for executing 
a security eject attempt. But the 
cassette is completely filled.
• Allow at least 1 empty position in 
the cassette.
• If the cassette is not filled com-
pletely, then clean the sensor sur-
faces.
• Adjust sample detection sensor 
threshold (Bruker service only).
103.310.063 Sample Detect at 
magnet failed.
A sample was not detected in 
the magazine for the requested 
eject.
Check if:
• There is really no sample in the 
magnet (or the sample down 
detection sensor threshold is set 
up not correctly).
• The lift air flow is strong enough to 
eject the sample back into the 
magazine.
• The sample detection sensors 
work well.
103.310.094
103.310.096
117.310.058
117.310.038
117.310.124
Missing empty posi-
tion for eject.
The position where the sample 
from the magnet should be 
ejected to is occupied. Allow at 
least 1 empty position in the 
cassette.
• Check if there is really a sample.
• Clean sensor surfaces.
• Adjust sample detection threshold 
(Bruker service only).
103.310.097 Sample from mag-
net was placed in 
position XX.
The source position of the sam-
ple in the magnet was 
unknown.
103.310.099
103.310.100
103.310.101
Source position not 
empty.
The position in the cassette 
from where the sample was 
inserted into the magnet is 
occupied.
• Do not fill the position from the 
sample down into the magnet.
103.310.103 Source position not 
empty, no other 
empty position 
found.
The position in the cassette 
from where the sample was 
inserted into the magnet is 
occupied and there is no other 
empty position available.
• Do not fill the position from the 
sample down into the magnet.
103.310.131
103.310.134
103.310.136
Missing empty posi-
tion.
There is no empty position in 
the cassette for the operation 
requested.
• Allow at least 1 empty position in 
the cassette. 
• If the cassette is not filled com-
pletely, then clean the sensor sur-
faces. 
• Adjust sample detection threshold 
(Bruker service only).