A-14
Appendix A Scanner
®
2000 microEFM
Measurement Canada Seal Kit
Measurement Canada has approved the use of the Scanner 2000 for custody transfer applications when it is
installed in accordance with the conguration and sealing provisions cited in Measurement Canada Approval
No. AG-0557C.
Measurement Canada compliance requires both the installation of a lead seal on the device and the enabling
of a custody transfer device seal in software that effectively prevents the user from changing the device
conguration without breaking the seal. For best results, congure the Scanner 2000 using ModWorX Pro
software prior to installing the seal kit. See the ModWorX Pro User Manual, Part No. 9A-30165025, for
information on conguring the device.
Seal kit components are packaged in a small plastic bag for shipment with Measurement Canada approved
devices. The seal kit includes the following components:
• a double-strand seal wire with a lead seal attached to one end
• an Allen head screw drilled to accept a seal wire
• a small Allen wrench
• an S-shape metal bracket drilled to accept a seal wire
• a circuit board jumper (for activating the device conguration lock)
Measurement Canada approved units can be identied by a secondary tag containing unit specications and
the Measurement Canada approval number. The tag is afxed to the outside of the Scanner 2000 enclosure
prior to shipment.
Seal Kit Installation
To install the Measurement Canada seal kit, perform the following steps.
1. Remove the cover from the Scanner 2000.
2. Remove the seal kit components from the plastic bag.
3. Remove the two screws from the Scanner 2000 switchplate and set aside.
4. Install the seal kit jumper as follows.
a. Pull the switchplate and circuit board assembly forward to access the back side of the circuit board.
b. Locate the J2 receptacle labeled “SWITCH” and insert the jumper into the header. (See Figure A.19,
page A-15)