EasyManua.ls Logo

Cameron Scanner 2000 microEFM - Installation Procedure-Remote Mount to Orifice Meter or Cone Meter

Cameron Scanner 2000 microEFM
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
Scanner
®
2000 microEFM Section 2
1. Verify that the meter is properly installed in the ow line (per manufacturers instructions).
2. Bolt a ange-by-ange 5-valve manifold (as recommended by Cameron) to the Scanner 2000 MVT sen-
sor.
a. Locate the H and L markings on the integral MVT sensor body and position the MVT/manifold as-
sembly so that the upstream side of the ow line can easily be connected to the sensors “High” port
and the downstream side of the ow line can be connected to the sensors “Low” port. The Scanner
2000 enclosure can be rotated to face the desired direction.
b. Position the manifold so that all valves are accessible from the front of the instrument.
3. Connect the Scanner 2000 and manifold assembly to the differential pressure meter. Hardware require-
ments will vary, depending upon the installation conguration. However, minimally, an adapter is re-
quired that can span between the threaded pressure tap/orice ange connector and the non-threaded
manifold. This adapter can be a one-piece stabilizer (often preferred for added strength and stability) or a
short heavy wall pipe nipple attached to a futbol ange (available from Cameron). Use a suitable com-
pound or tape on all threaded process connections.
CAUTION Do not use Teon
®
tape on the threads of the union, adapter, or pipe plugs. Use of
Teon
®
tape will void the explosion-proof rating of the instrument.
4. Install the RTD assembly in the thermowell. Route the RTD assembly cable through the conduit opening
in the top of the Scanner 2000 to connect to the main circuit board. A wiring diagram for the RTD assem-
bly is provided in Figure 3.5, page 66. For hazardous areas, review Hazardous Area Installations, page 27.
5. Route any additional inputs/outputs or COM connections, etc. through the conduit opening in the top of
the Scanner 2000. For hazardous areas, review Hazardous Area Installations, page 27.
6. Perform a manifold leak test as described on page 55.
7. Verify the zero offset, if required (and other calibration points, if desired). See the ModWorX™ Pro Soft-
ware User Manual, Part No. 9A-30165025, for complete instructions. See also Zero Offset (Static Pres-
sure or Differential Pressure), page 56, Static Pressure Calibration and Verication, page 56, and Differential
Pressure Calibration and Verication, page 57.
CAUTION Do not put the Scanner into operation until the valves are positioned properly so that
pressure is supplied to both sides of the MVT. For instructions on proper valve positions,
see Placing the Scanner into Operation, page 58.
Installation Procedure—Remote Mount to Orice Meter or Cone Meter
A Scanner 2000 can be mounted remotely and connected to an orice meter or cone meter with tubing for gas
measurement. The setup of the meter run and plumbing congurations can vary widely, depending upon the
challenges existing on location. Figure 2.6 shows a typical remote-mount installation.
WARNING HAZARDOUS AREA USE. The Scanner 2000 is certied for hazardous area use only when
installed in accordance with applicable standards and local wiring practices. Carefully
review Hazardous Area Installations, page 27, to determine specic installation require-
ments (cable glands, conduit seals, signal cable, RTD, etc.).

Table of Contents

Related product manuals