THS – FI022GB2K8v3 --- V - MAINTENANCE
75
Remote Possible settings Standard setting Parameter
Bin.Full al.
LF
ON / OFF OFF
Global
Description Models
Enabling of “BIN_FULL” alarm
If the input remains active for more than 5 seconds, the malfunction signal is activated.
The alarm is reset by pressing the
key.
All models
Remote Possible settings Standard setting Parameter
Bin Absent alarm
LA
ON / OFF OFF
Global
Description Models
Enabling of “BIN_ABS” alarm
If the input remains active for more than 5 seconds, the malfunction signal is activated.
The alarm is reset by pressing the
key.
All models
Remote Possible settings Standard setting Parameter
Air pressure al.
AP
ON / OFF OFF
Global
Description Models
Alarm in case of insufficient air pressure
OFF: LOW_PRESS input ineffective; ON: if the LOW_PRESS input is active the system goes into “Low Pressure”
fault mode.
All models
Remote Possible settings Standard setting Parameter
Motor alarm
MA
ON / OFF -
Global
Description Models
Motor overheating alarm
OFF: alarm disabled (configuration without motor driving or with MDL card)
ON: Version with MDT card
: only alarm if the thermic protection sensor, connected to J25
of MDT card, is active
All models,
except THS/G
Remote Possible settings Standard setting Parameter
Ejection
EJ
ON / OFF ON
Global
Description Models
Ejection
Enabling/disabling of ejection, used only for servicing purposes.
All models
Remote Possible settings Standard setting Parameter
Ejec. queue reset
QR
ON / OFF ON
Global
Description Models
Reset of the ejection queue
OFF: ejection queue is never reset. If the belt stops and a contaminated package is placed between the antenna and
the ejector, the package is ejected when the belt is started again;
ON: if there is an MDL card or an encoder, the ejection queue is reset when the belt stops, whether by command or
as the result of a fault.
All models,
except THS/G
Remote Possible settings Standard setting Parameter
Fault relay
FR
ON / OFF OFF
Global
Description Models
Enabling of fault relay
If FR=ON, and in case of fault the “fault relay” and flashing indicator, connected to J22 pin 8 and 9, are activated.
All models
Remote Possible settings Standard setting Parameter
DTR protocol
DT
N / H N
Global
Description Models
DTR protocol
Use of DTR line to communicate with printer.
N = line not used; H = line used, handshake mode, high active line.
The metal detector stop sending data to the serial port, in case the line is not pulled up.
All models