Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 11
For your safety and that of your patients
Connected devices
Networking and connection to other devices
When combining Dräger devices with other electrical devices, the owner must ensure that the resulting
system meets the requirements of the following standards:
– IEC 60601-1 (EN 60601-1)
Medical electrical equipment
Part 1: General requirements for safety
– IEC 60601-1-2 (EN 60601-1-2)
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic compatibility; Requirements and tests
Strictly follow the Assembly Instructions and Instructions for Use for each connected device.
Connection to hospital network
Many medical devices manufactured by Dräger use networks to transmit patient data in real-time and to
notify clinical users of alarm conditions. Hospitals should refer to IEC 80001-1 before attempting to
connect such medical devices to their IT networks.
IEC 80001-1 requires manufacturers (such as Dräger) to make technical documentation in support of such
network connections available upon request. Contact your Dräger representative for that information or to
facilitate negotiation of an IEC 80001-1 Responsibility Agreement for additional support from Dräger.
WARNING
Risk of electric shock and of device malfunction
Any connected devices or device combinations not complying with the requirements mentioned
in these instructions for use can compromise the correct functioning of the medical device and
leads to an electric shock. Before operating the medical device, strictly comply with the
instructions for use of all connected devices and device combinations.