Problem solving
238 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
M300 messages
The following table lists various messages that may display on the M300.
Priority Message Description Action
None ƣ All alarms paused The alarm function of the
M300 device is temporarily
disabled. The message is
followed by minutes and
seconds which represent a
count-down timer.
Informational message,
no action required.
None ě Audio paused The audible alarm signal for
the M300 device is silenced
for 2 minutes. The message
clears after the alarm si-
lence period times out.The
message is followed by
minutes and seconds which
represent a count-down
timer.
Informational message,
no action required.
None
Default patient settings re-
stored
User settings were restored
to factory defaults.
Informational message,
no action required.
None
Discharged The patient is discharged
(see page 110).
Informational message,
no action required.
None High temperature reset Reset due to the internal
temperature exceeding the
acceptable limit.
Informational message,
no action required.
None
HR alarms off
Alarms related to the heart
rate parameter are dis-
abled.
Informational message,
no action required.
None
HR,ASY,VF off
1
– Arrhythmia monitoring
is disabled.
–HR alarms are disabled
– The ASY/VF alarms fea-
ture is set to Follow HR
(not supported by the
M300).
Informational message,
no action required.
! Low battery The M300 battery charge is
low, with approximately
30 minutes of runtime re-
maining.
Recharge battery.