Recordings/reports
200 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Actions that affect recordings
M300 offline recordings
While the M300 is offline or the R50N recorder is not available, the last five recordings (automatic or
manual) are stored on the M300 when the Store setting is applied. When the connection is restored, those
recordings are sent to the R50N recorder.
Action Effect on pending recordings Effect on recordings in progress
Patient Discharge Pending recordings are discarded Recording is canceled
Edit patient’s demographic
data
No effect No effect
Patient monitoring put in
Standby mode
No effect Recording is canceled
Configuration change of
Main screen resulting in re-
moval of a patient
Patient’s pending recordings are
canceled
Recording is canceled
Change in scale settings or
parameter order (parame-
ters might become avail-
able/unavailable due to re-
stored/missing signals)
No effect Recordings will continue but wave-
form and parameter data for any re-
moved parameter appear as blanks.
Added parameters are not included
on a recording currently in progress.
Recorder assignment
change
No effect. Additional recordings are
printed on the new recorder.
No effect on recordings.
Network time/date change No effect on active or pending re-
cordings. All new recordings in
progress reflect the new date/time.
Recordings reflect the change im-
mediately (as will all future manual
recordings).
R50N recorder becomes
unavailable, runs out of pa-
per, or jams
No effect Recording is not lost, but routed to
another assigned recorder, or
stored for printing when the record-
er becomes available.
Remote device requesting
a recording of a local pa-
tient goes offline
Deletes all recordings associated
with this device
Recording is canceled
Patient disconnect for
transport
Cancels all pending recordings for
that bed
Recording is canceled