For your safety and that of your patients
16 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Site of operation
Only use devices (monitors, chargers, accessories, etc.) in areas that meet the environmental
requirements outlined in the Technical Data chapter.
WARNING
Do not connect connectors with an ESD warning symbol and do not touch their pins
without implementing ESD protective measures. Such protective measures may
include antistatic clothing and shoes, touching a potential equalization pin before
and during connection of the pins, or using electrically insulating and antistatic
gloves. All users concerned must be instructed in these ESD protective measures.
WARNING
Do not operate the device in the following areas: magnetic resonance imaging (MRI)
environments, aircraft, ambulance, home, or hyperbaric chambers.
WARNING
To avoid interfering with device operation, do not operate devices (monitors, accessories, etc.)
within 10 m (33 feet) of equipment that emits microwave or other high-frequency emissions.
WARNING
To minimize the risk of patient strangulation, carefully position and secure sensor cables. Also
position the sensor cables to minimize inductive loops.
WARNING
When placing the Infinity M300 and M300+, make sure adequate airflow exists. Do not cover the
device with blankets.
CAUTION
After extended exposure in a cold environment, acclimate the device carefully so that condensation does
not form on the electronic parts and damage the device.
CAUTION
To avoid damaging the touch-sensitive screen, do not allow sharp instruments to touch the front panel
of the devices.
WARNING
To avoid short-circuiting and otherwise damaging the device, Dräger recommends that no fluids come
in contact with the ICS devices when they are connected to a power socket. If fluids are accidentally
spilled on the equipment, remove the affected device from service as soon as possible and have tech-
nical personnel verify that patient safety is not compromised.