SpO2 and pulse rate monitoring with Nellcor and M300
190 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Accessing the SpO2 setup page
1 Open the BedView screen by selecting the viewport of the telemetry patient on the ICS Main screen.
2 Select the Telemetry/ Parameters button.
3 Select the Layout/ Parameters tab (if not already selected).
4 Select the SpO2 tab.
SpO2 parameter setup functions
All SpO2 parameter setup functions take place in the ICS SpO2 setup page (see "Accessing the SpO2
setup page" on page 190).
To configure the alarm functions, see page "M300 alarm ranges and defaults" on page 134.
Selection Available settings Description
SpO2 monitoring On, Off (default) Turns SpO2 monitoring on or off.
Bar graph On (default), Off Displays a bar graph pulsating at pulse rate with
the amplitude being proportional to the pulse
strength. Displays only on the M300 not the ICS.
Color Red, white (default), yellow,
green, light blue, blue, purple,
orange
Determines the color of the waveforms and the
parameter labels and values.