Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 95
Charging the M300 battery
Using the Bedside Charger
The following illustration shows a Bedside Charger, which is used to charge an individual M300. When the
green LED (A) of the Bedside Charger is illuminated, the Bedside Charger is connected to power and is
ready to charge an M300.
A Power LED – lights up green when the Bedside Charger is plugged in
B Power connector/plug
Charging M300 at the bedside
To place the M300 into the Bedside Charger
1 Plug the power cable (A) into a power outlet.
2 Place the bottom of the M300 in the Bedside Charger, as shown in the following illustration.
425
CAUTION
When docked on the bedside charger, make sure that the Infinity M300 is not placed under bed covers
and/or sheets. In addition, do not place the M300 and bedside charger inside the patient’s pocket or car-
rying pouch for the following reasons:
the M300 requires air circulation to dissipate heat resulting from the charging battery
the M300 will suspend battery charging if it detects that heat is not being dissipated due to a lack of
adequate air circulation
the M300 alarm volume can be reduced if located under bed covers or sheets. This can potentially
prevent the M300 audio (including alarms or fault mitigation) from being heard.
WARNING
Place the M300 and the Bedside Charger so they do not fall on a patient thus causing injuries.
Make sure that no accessory cables accidentally become entangled around the neck of a patient.