System configuration
222 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Layout/ Parameters > SpO2 page
The SpO2 page enables you to define default system-wide SpO2 monitoring settings, and to adjust the
settings displayed in the following table.
Layout/ Parameters > ST page
The ST tab is inaccessible until the ST locked option is enabled. For additional information, contact your
technical personnel of your hospital.
The ST setup page is where you can customize ST monitoring and define system-wide default settings
for the ST parameter boxes. In addition, the ST page enables you to set up the ST units of measure and
ST measurement point.
ST alarm limits default settings must be configured in Telemetry defaults > Alarms > Limits tab (see
page 209). For additional information, see "Pausing alarms for the M300 from a remote network device"
on page 127.
Following admission, ST parameter order can be changed on each patient-specific BedView ST setup
page.
The following table displays the available selections within the ST system default setup page.
For additional information about ST monitoring, see page 159.
Selection Settings Description
SpO2 monitoring On, Off (default) Enable or disable SpO2 monitoring.
SpO2 Monitoring can be enabled for local te-
lemetry patients only.
This option appears unavailable for selection
when SpO2 is unavailable.
Bar graph On (default), Off When set to On, the M300 displays a pulse bar
in its SpO2 parameter box. The bar graph is
not displayed in the ICS parameter box.
Color Red, White (default), Yellow,
Green, Light Blue, Blue, Purple,
Orange
Select the colored box and then make the ap-
propriate color selection for SpO2 waveforms
and parameter box values, units of measure,
and labels.