Problem solving
232 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
M300 – ECG alarm messages
The following table lists ECG alarm messages that may display on the M300.
Priority Message Problem Solution
! ECG artifact
1
Patient movement
(shivering, tremors)
Bad electrode contact
Excessive signal
noise interference
from auxiliary equip-
ment
Check the electrodes and
reapply if necessary.
Make sure that the pa-
tient’s skin is properly
prepped.
Isolate the patient from
auxiliary equipment, if
possible.
!
!!
!!!
ECG leads off
1
Lead-off condition detect-
ed due to:
Broken cable(s)
Disconnected ECG
lead wires
Loose lead wire(s)
Wrong lead selected
Dried out electrode gel
Replace defective ca-
ble(s).
Reapply gel on reusable
electrodes.
Replace dried electrodes
with fresh ones with good
gel.
Select another ECG lead
for processing.
If monitoring augmented
leads, verify that the num-
ber of selected leads in
the ECG setup page is
correct.
Check cable(s) and con-
nection(s)
Replace cable(s) if neces-
sary
!! HR > (alarm limit)
HR < (alarm limit)
The parameter value is
above/below the set
upper /lower alarm limits.
Check the patient and
treat if necessary.
Change the alarm limits.
1
In the parameter box the parameter value is replaced by ***