ECG, arrhythmia, and ST segment monitoring with M300
154 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Selecting arrhythmia leads
Appropriate lead selection is essential for accurate arrhythmia monitoring. It is ideal to assign the two best
ECG leads as the arrhythmia monitoring leads. The following two options are available:
– ECG 1 (single channel option)
– dedicates processing to the lead selected as the top waveform dis-
played.
– ECG 1 & 2 (Dual channel option)
– determines the heart rate and arrhythmia based on the leads se-
lected as top two waveforms.
To select default arrhythmia leads
1 Select the Telemetry defaults button in the ICS Main screen.
If the selection does not appear, the Telemetry option is not enabled.
2 Select the Layout/ Parameters tab (if not already selected).
3 Select the Layout tab.
4 Select the arrow next to QRS/ARR processing to choose the appropriate channel(s).
To select arrhythmia leads for an individual M300
1 Open the BedView page by selecting the parameter box of the telemetry patient in the ICS Main
screen.
2 Select the Telemetry/ Parameters button.
3 Select the Layout/ Parameters tab (if not already selected).
4 Select the Layout tab (if not already selected).
5 Select the arrow next to QRS/ARR processing to choose the appropriate channel(s).