Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 17
For your safety and that of your patients
Defibrillator precautions
The M300 and Bedside Charger are protected against high-frequency interference from defibrillators and
against 50-Hz and 60-Hz power line interference.
CAUTION
To prevent burns and electric shock due to rerouting of electrical current through electrodes, do not po-
sition the defibrillator pads near any electrodes or sensors.
CAUTION
Only defibrillate across the chest.
CAUTION
Using ECG electrodes and cables specified by Dräger protects the device from damage during defibril-
lation and reduces noise and other interference on the ECG waveform.
CAUTION
Remove all non-defibrillation-proof applied accessories prior to defibrillation. Failure to remove non-de-
fibrillation-proof applied accessories could potentially be harmful to both the patient and operator, as well
as other medical devices connected to the patient.