ECG, arrhythmia, and ST segment monitoring with M300
148 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
ECG parameter setup functions
All ECG parameter setup functions take place in the ECG setup page (see page 147).
To configure the alarm functions, refer to the Instructions for use Infinity CentralStation Wide, Software
VG4.0 or later.
Selection Available settings Description
TruST 12-lead On, Off (default) Enables/disables TruST moni-
toring (see page 162).
(Selection only appears when the TruST locked option is enabled.)
Cable type 3-, 5-, and 6-lead (default) wires Selects the lead wire set.
(TruST is available only with a 6-lead wire set.)
Pacer detection On (default)
Depending on the bedside monitor software
version, the ICS will display a Pacer ON ban-
ner for both Pacer ON and Pacer Fusion pac-
er detection settings. Refer to the bedside
monitor’s display to identify which pacer de-
tection mode is selected.
Off (the message Pacer off appears in the
waveform area)
Fusion – is available only if Pacer fusion is
enabled; see page 203 (the message Pacer
fusion appears in the waveform area)
Determines whether pacer im-
pulses are detected. See
"Pacer Fusion mode"
on page 151 for precautions
before you start this mode.
Color Red, white, yellow, green (default), light blue,
blue, purple, orange
Determines the color of the
waveforms, and the parameter
labels and values.