Alarms
Parameter box flashing rate
The parameter box will flash at a 50% duty cycle for high- and medium-grade priority alarms, as follows:
High-priority alarm = 1.4 to 2.8 Hz
Medium-priority alarm = 0.4 to 0.8 Hz
The parameter box for Low-priority alarms will remain solid (no flashing).
The M300 transceiver sends audio pause information to the network so the ICS can display the "Audio
paused" status and counter.
When the monitor is in an ’All alarms paused’ state, a banner displaying “All alarms paused mm:ss
appears in black text within a yellow background, (where mm:ss represents a countdown timer in
minutes:seconds).
The priority of the alarm banners are, as follows:
When the HR limits and Arrhythmia are disabled, a "ĉ HR, ASY, VF off" banner will appear as white text
on a red background.
Note that the message will rotate through any other active messages (such as alarms, or status
messages).
Note that the "ĉ HR, ASY, VF off" message will rotate through any other active M300 messages (such
as alarms or status messages).
When the HR limits are disabled and Arrhythmia is enabled, a "ĉ HR alarms off" banner will appear as
black text on a yellow background.
Note that this banner message will also rotate through any other active M300 messages (such as alarms
or status messages).
All alarms paused mm:ss
Audio paused mm:ss
HR, ASY, VF off