M300 and M300+ series – VG3.0 257
Reprocessing
Reprocessing of patient-specific accessories
Categorization of accessories
Reprocessing of the categories ECG monolead, SpO2 intermediate cable and SpO2
sensor
Surface disinfection with cleaning
Prerequisites:
The surface disinfectant has been prepared in accordance with the manufacturer's instructions.
The manufacturer's instructions, e.g., regarding shelf life or application conditions, are observed.
An uncontaminated, lint-free cloth soaked in surface disinfectant is used for the cleaning surface dis-
infection.
Category Classification Part number Description of the
procedure
ECG monolead Non-Critical MS14555
MS14556
MS14559
MS14560
MS14683
MS14682
See "Surface disinfection
with cleaning" on page 257.
SpO2 intermediate cable Non-Critical MS18683
MS17330
SpO2 sensor Non-Critical MS13235
Pouch M300+ disposable 100 pcs Non-Critical MS34194
Pouch M300+ waterproof 10 pcs Non-Critical MS33867
Pouch M300 waterproof new tag Non-Critical MS32401
Tele transmitter pouch disposable
100 pcs
Non-Critical MS22905
Surface disinfectant Manufacturer Concentration Contact time
Oxycide Ecolab USA 2.3 % 5 min
WARNING
Risk due to penetrating liquid.