System overview
54 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Battery safety information
Observe the following warnings and cautions for the swappable battery.
Swappable Battery run time on the M300+
WARNING
Risk if the swappable battery is not replaced properly.
If the swappable battery is not reattached or serviced and replaced properly, short circuits and
high temperatures leading to explosion or fire may occur.
WARNING
Risk if the swappable battery contacts are wet.
If the battery is detached or removed for service and the electrical contacts are moistened with
disinfectant, immediately reattaching or reinstalling the battery while the contacts are still wet
could lead to corrosion and contact failure. Dry the contacts before reattaching the battery.
WARNING
Do not crush, drop, or puncture the swappable battery as this could lead to internal damage and
internal short circuits which may not be visible externally.
WARNING
Do not incinerate or expose the swappable battery to high temperatures.
WARNING
Discontinue use of defective or damaged batteries, such as swappable batteries that do not lock
into place on the M300+.
CAUTION
Use M300+ compatible Dräger swappable battery only.
CAUTION
Dispose of swappable batteries in compliance with your institution’s disposal instructions.
M300+ with swappable battery
Typical Battery Run Time Load Conditions
802.11 b/g/n 2.4 GH
Z: 22 hours
802.11 a/n 5 GHz: 22 hours
– ECG only
– Display off
–Audio off