Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 291
Technical specifications
EMC declaration for M300/M300+
General information
This device was tested for electromagnetic compatibility using accessories from the list of accessories.
Other accessories may only be used if they do not compromise the electromagnetic compatibility. The use
of non-compliant accessories may result in increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of other devices only if Dräger has approved this device
arrangement. If no approval has been given by Dräger, it must be ensured that this device functions
correctly in the desired arrangement before use. The instructions for use for the other devices must be
followed.
Electromagnetic environment
This device may only be used in environments specified in section "Overview of the M300/M300+"
on page 43.
WARNING
Risk due to electromagnetic disturbance
Wireless communication devices (e.g., cellular phones) and medical electrical equipment (e.g.,
defibrillators, electrosurgical devices) emit electromagnetic radiation. When such devices are
operated too close to this device or its cables, the functional integrity of this device may be
compromised by electromagnetic disturbances. As a result, the patient could be put at risk.
Maintain a distance of at least 0.3 m (1.0 ft) between this device and wireless communication
devices, to ensure that the essential performance of this device is fulfilled.
Maintain an adequate distance between this device and other medical electrical equipment.
Emissions Compliance
Radiated emissions Class A, group 1 (30 MHz to 1 GHz)
Conducted emissions Class A, group 1 (150 kHz to 30 MHz)
NOTE
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required), this equipment might not offer adequate protection to radio-frequency communication ser-
vices. The user might need to take mitigation measures, such as relocating or re-orienting the equip-
ment.