292 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Recommended separation distances from wireless communication devices
To ensure that the functional integrity of this device is maintained, there must be a separation distance of
at least 1.0 m (3.3 ft) between this device and wireless communication devices.
Operating characteristics
Classification
Immunity against
IEC 60601-1-2 4th edition
Test level and required electromagnetic
environment
Electrostatic discharge (ESD) (IEC 61000-4-2) Contact discharge: ±8 kV
Air discharge: ±15 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
30 A/m
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V
RMS, ISM bands: 6 VRMS
Electromagnetic fields in the vicinity of wireless
communication devices
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m, with various pulse modulations and
immunity test levels.
Classification Medical Device Europe Class IIb