Reprocessing
250 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Before reprocessing
Observe before disassembly
1 Turn off the M300/M300+ Telemetry Device:
a. Simultaneously press and hold both arrow keys.
b. Select the up arrow to highlight Yes.
c. Press the Menu button and confirm the shutdown.
2 Unplug the M300 Bedside and Central Charger:
a. If docked, remove the M300 telemetry device.
b. Disconnect the power plug.
3 Unplug the M300+ Battery Charger:
a. Remove all batteries from the charger.
b. Disconnect the power plug.
WARNING
A risk of an electric shock might occur due to penetrating liquid. Follow the instructions below
to disconnect power before reprocessing.