EasyManua.ls Logo

Dräger Infinity M300 series - M300+ Swappable Battery

Dräger Infinity M300 series
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 53
System overview
M300+ swappable battery
The new Infinity M300+ operates with a swappable, externally rechargeable battery with the following
specifications:
Battery recharging time of approximately 3.5 hours
Battery swaps retain clinical data and current alarm settings for approximately two minutes
Battery swaps do not require tools
Battery provides a four LED charge level indicator when the battery is detached
The complete unit, with the swappable battery pack and the battery retainer attached to the
monitor, is rated IPX4, protected against splashing water
WARNING
Do not try to power up the M300+ without a battery installed. Attempting to power up the M300+
without a battery installed will result in erasing any stored patient information from the M300+.
NOTE
IPX ratings:
Main unit with battery attached is IPX 4, protected against splashing water.
Main unit without battery attached is IPX 7, protected against temporary immersion.
Swappable battery alone, unattached, is IPX4, protected against splashing water
WARNING
During battery replacement, there is no data monitoring and no transmission of data to the ICS.
However, clinical data and current alarm settings are retained in the M300+ for a maximum of two
minutes. Exceeding this time limit will result in loss of clinical data and alarm settings.

Table of Contents

Other manuals for Dräger Infinity M300 series

Related product manuals