Cybersecurity
32 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Network support matrix
The following matrix identifies the interaction of Dräger devices on the Infinity network when the operation
mode is secure and legacy:
Product Security
Mode
ICS
VG4
ICS
VG4
ICS
VG3 and
below
M300 (+)
VG3
M300 (+)
VG3
M300 (+)
VG2.4
IACS/
M540
VG7.1.1
and
below
Delta
2)
VF10
and
below
Security
Mode
Secure Legacy Legacy Secure Legacy Legacy Legacy Legacy Secure
Action Admit Admit Remote
View
Remote
View
Remote
View
Remote
View
Remote
View
Remote
View
Transfer
M300/
Bedside
ICS VG4 Secure - - - Y Partial
1)
1) In remote view with Legacy mode, the waveforms and patient parameters can be viewed. Remote control (alarm settings,
arrhythmia settings, alarm silence, pause, and record) are not supported.
Partial
1)
YY-
ICS VG4 Legacy - - - Partial
1)
YYYY-
ICS VG3
and below
Legacy - - - Partial
1)
YYYY-
M300 (+) VG3 SecureYNN-----Y
M300 (+) VG3 Legacy N Y Y - - - - - N
M300 (+)
VG2.4
Legacy N Y Y - - - - - N
M540 VG7.1.1
and below
Legacy Y Y Y - - - - - -
IACS
VG7.1.1
and below
Legacy Y Y Y Partial
1)
YYYY-
Delta
2)
VF10
and below
2) DSCP (Quality of Service) must be set to 0 for Delta devices.
Legacy Y Y Y Partial
1)
YYYYN