Problem solving
234 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
M300 – ST alarm messages
The following table lists ST alarm messages that may display on the M300.
Priority Message Problem Solution
!! Cannot analyze ST The algorithm cannot
determine ST values
due to artifact, the ab-
sence of normal beats,
or invalid leads.
– Perform a relearn
(see page 169).
– Check electrodes; reapply if
necessary.
– Make sure the patient’s
skin is properly prepared.
– Isolate the patient from
auxiliary equipment if pos-
sible.
– Inspect and replace defec-
tive cable(s) and wire(s).
– Reapply gel on reusable
electrode(s).
– Reapply the electrode(s).
Make sure the patient’s
skin is properly prepared.
– If a lead or electrode can-
not be replaced, select an-
other ST lead for
processing.