Reprocessing
258 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Cleaning
1 Wipe off obvious soiling with a disposable cloth soaked in surface disinfectant.
2 Dispose of the cloth.
3 Wipe all surfaces with a new disposable cloth soaked in surface disinfectant. After that, there must no
longer be any soiling visible.
Surface disinfection
4 Take a new disposable cloth soaked in surface disinfectant. Wipe cleaned surfaces again until all sur-
faces to be disinfected are visibly wet.
5 Wait 15 minutes for the surface disinfectant contact time.
6 At the end of the contact time, moisten a new, uncontaminated and lint-free cloth with water (at least
drinking water quality).
7 Wipe all surfaces until no remains of the surface disinfectant, such as foam residues or streaks, are
visible.
8 Wait until the surfaces are dry.
9 Check the surfaces for visible damage and, if necessary, replace the product.
WARNING
Risk of electric shock and device malfunction.
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.
CAUTION
To avoid damaging the device, do not use sharp tools or abrasives. Never immerse electrical connectors
in water or other liquids.
CAUTION
Do not autoclave accessories.
CAUTION
Do not use excessive pressure or flex patient cables unnecessarily when cleaning. Excessive pressure
can damage the patient cables.