Instructions for use – Infinity
®
M300 and M300+ series – VG3.0 237
Problem solving
!
!!
!!!
SpO2 sensor off
Nothing has been detect-
ed between the sensor's
light source and detector
for 5 seconds.
Ensure sensor is properly
connected to patient's
finger
!
SpO2 light blocked
Insufficient light is reach-
ing the sensor's light de-
tector for a valid mea-
surement.
Ensure that light detector
side of the sensor is not
blocked and only patient
finger is placed between
light source and detector.
!
SpO2 no measurement
30 seconds have
elapsed since a valid
measurement was com-
puted and no other tech-
nical condition exists.
Ensure that sensor is prop-
erly attached to patient and
not moving. Reposition the
sensor or use a different
measurement site.
!
SpO2 weak signal
Pulsations are not of suf-
ficient amplitude to com-
pute a valid measure-
ment
Reposition the sensor or
use a different measure-
ment site.
!
SpO2 Non-Masimo sensor
Nellcor sensor used with
monitor SW configured
for Masimo sensor
Ensure that Masimo sensor
is plugged into M300 trans-
ceiver.
!
SpO2 Non-Nellcor sensor
Masimo sensor used with
monitor SW configured
for Nellcor sensor
Ensure that Nellcor sensor
is plugged into M300 trans-
ceiver.
!
SpO2 regulation error
Input from the sensor can
not be maintained at
proper levels for a valid
measurement to take
place
Ensure that sensor is prop-
erly attached to patient and
not moving.
1
In the parameter box, the parameter value is replaced by +++
2
In the parameter box, the parameter value is replaced by - - -
3
In the parameter box, the parameter value is replaced by ***
Priority Message Problem Solution