EasyManua.ls Logo

Dräger Infinity M300 series - Page 14

Dräger Infinity M300 series
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
For your safety and that of your patients
14 Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
WARNING
Inaccurate SpO2 measurements may result under the following conditions:
Elevated levels of methemoglobin
Elevated levels of total bilirubin
Excessive patient motion
Severe anemia
Low arterial perfusion
WARNING
Risk of electric shock and device malfunction.
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
WARNING
Pacemaker considerations
False low-rate alarms can result under the following conditions:
Asynchronous pacemaker with large amplitude pace pulses with no overshoot and at low
heart rate (30 bpm)
False high-rate alarms can result under the following condition:
Asynchronous pacemaker with large pace pulse tails and at low heart rate (30 bpm)
CAUTION
Signs of wear, such as cracks, deformation, discoloration, or peeling, may occur with reprocessed prod-
ucts.
Check the products for signs of wear and replace them if necessary.
CAUTION
To avoid damaging the device, do not use sharp tools or abrasives. Never immerse electrical connectors
in water or other liquids.
CAUTION
Do not autoclave accessories.
WARNING
Risk of explosion and of chemical burns
Improper handling of batteries can result in explosions and chemical burns.
Do not throw batteries into fire. Do not force batteries open.

Table of Contents

Other manuals for Dräger Infinity M300 series

Related product manuals