Instructions for use Perseus A500 SW 2.0n 145
Alarms
inAgent high ---
After two successive
respiratory phases
and 15 seconds
or
after 30 seconds if no
respiratory phases are
detected
>165 seconds later
Inspiratory xMAC high ---
insp. MAC
≥
3
for more than
180 seconds
longer than
30 seconds:
insp. MAC
≥
3
and
exp. MAC
≥
2.5
or
insp. MAC
≥
5
xMAC low 0 to 60 s >60 s ---
Apnea (no CO
2)
Apnea (no flow)
Apnea (no pressure)
Apnea
---
at the latest after
20 seconds
(for RR
≥
6)
or
At the latest after
35 seconds
(for RR <6)
or
at the latest after
65 seconds for the
Pause, Manual /
Spontaneous and Ext.
FGO modes
15 seconds later
(for RR
≥
6)
or
30 seconds later
(for RR <6)
Apnea Ventilation
At the latest after 20 seconds
(15 seconds for RRmin
≥
4)
(configurable, see page 155)
---
No CO
2 detected >60 s --- ---
Inspiratory tidal volume high
Tidal volume not achieved
---
After 3 successive
breaths
---
Alarm
Priority
Low Medium High