Configuration
158 Instructions for use Perseus A500 SW 2.0n
Vertical tab "Vent. 2"
Headline/
Parameter
Setting range
Description
Start settings for ventilation Specify start settings for the venti-
lation.
Pmax
[mbar]; [hPa];
[cmH
2O]
7 to 80
40
7 to 80
30
7 to 80
25
Pinsp
[mbar]; [hPa];
[cmH
2O]
3 to 80
15
3 to 80
15
3 to 80
15
Psupp
[mbar]; [hPa];
[cmH
2O]
Off;
1 to 80
10
Off;
1 to 80
10
Off;
1 to 80
10
Insp. term.
[%PIF]
5 to 80
25
5 to 80
25
5 to 80
25
PEEP
[mbar]; [hPa];
[cmH
2O]
Off;
2 to 35
3
Off;
2 to 35
3
Off;
2 to 35
3
Slope
[s]
0 to 2
0.2
0 to 2
0.2
0 to 2
0.2
RRmin
[1/min]
Off;
3 to 25
6
Off;
3 to 25
10
Off;
3 to 25
15
I:E 1:50 to 50:1
1:2.0
1:50 to 50:1
1:2.0
1:50 to 50:1
1:1.0
%Tplat
[%]
0 to 60
20
0 to 60
20
0 to 60
20
Trigger
[L/min]
0.3 to 15
4.0
0.3 to 15
2.0
0.3 to 15
1.0
Sync. On; Off On; Off On; Off