Instructions for use Perseus A500 SW 2.0n 159
Configuration
Vertical tab "Proced. 1"
Start settings for APRV (optional)
Phigh
[mbar]; [hPa];
[cmH
2O]
3 to 80
15
3 to 80
15
3 to 80
15
Plow
[mbar]; [hPa];
[cmH
2O]
Off;
2 to 35
3
Off;
2 to 35
3
Off;
2 to 35
3
Tlow
[s]
0.2 to 10
4.0
0.2 to 10
2.0
0.2 to 10
1.0
Thigh
[s]
0.2 to 10
2.0
0.2 to 10
1.0
0.2 to 10
1.0
Slope
[s]
0 to 2
0.0
0 to 2
0.0
0 to 2
0.0
Headline/
Parameter
Setting range
Description
General settings
Default maneuver One-step recruitment; Multi-step recruitment;
Insp./Exp. hold
Defines the default maneuver.
Reminder
[min]
Off, 10 to 180 Off, 10 to 180 Off, 10 to 180 If the value is not set to Variable , a
reminder for a maneuver is given
after the first switch to a controlled
ventilation mode. Defines the time
after which, following the end of a
One-step recruitment or Multi-step
recruitment maneuver, a reminder
for a further maneuver is given.
Layout
Displayed parame-
ter
Cdyn; VT Defines which additional parame-
ter is displayed in the One-step
recruitment and Multi-step recruit-
ment dialogs.
Headline/
Parameter
Setting range
Description

Table of Contents